CH410-03 High Temperature LaBr3(Ce) Scintillation Detector (Energy spectrum measurement)

197

CH410-03 detector is a high temperature and vibration-resistant scintillation detector. It integrates high temperature LaBr3(Ce) scintillator, high temperature PMT and high linear voltage divider circuit. The unique design ensures the stable performance of the detector in high temperature vibration environment.

LaBr3(Ce) is a new type of scintillator. Compared with the traditional NaI(Tl) scintillator, LaBr3(Ce) has the advantages of better pulse amplitude resolution, higher detection efficiency for high energy ray, fast decay time and better temperature characteristic.

It is mainly used to measure formation elements in oil logging.


Specifications
Model
CH410-03 High Temperature LaBr3(Ce) Scintillation Detector (Energy spectrum measurement)
Detector diameter (mm)
Φ62
Detector length (mm)
258
Scintillator size (mm)
Φ49.5×101.6
Pulse amplitude resolution1) 25℃ Max.
6%
Stability2) (High temperature counting rate deviation) 175℃ Max.
3%
Diamagnetism3) (Geomagnetism) 25℃ Max..
1%
Vibration 4)
Random 5grms,50Hz~500Hz
Shock
100g@11ms
Operating temperature5) (℃)
25~+175
Storage temperature5) (℃)
30~+70

1) Test with137Cs;

2) Tested in natural gamma environment;

3) The detector is respectively placed in 6 directions of X+, X-, Y+, Y-, Z+ and Z-, and the output pulse amplitudes of the 6 directions are obtained through test in sequence, taking X+ as the reference, the other 5 directions is compared with the reference amplitude, and the maximum value of the change of the 5 directions relative to the reference amplitude is calculated;

4) Resonance frequency ≥500Hz;

5)Tempreature change rate during detector operation and storage≤1.8℃/min.